Enquire Now

Mastersizer 3000

MANUFACTURER Malvern (manufacturer's website)
CONTACT 1 David Race
Enquire about this item
SITE City Campus

Description

Measuring over the nanometer to millimeter particle size ranges, it packs exceptional performance into the smallest of footprints, bringing operator-independent measurements that every user can rely on. With excellent Malvern software, the total package allows for simple method development of research intensive quality.

Specification

  • PARTICLE SIZE

Suspensions, emulsions, dry powders (0.01-3500µm); 100 size classes

  • SIZE ACCURACY

>0.6%

  • SIZE PRECISION

>0.5% variation

  • SIZE REPRODUCTION

>1% variation

  • PRINCIPLE

Laser light scattering

  • OPTICS

Red light source (max. 4mW He-Ne, 632.8nm); blue light source (max. 10mW LED, 470nm); lens arrangement is Reverse Fourier (convergent beam); effective focal length (300mm)

  • DETECTOR

Log-spaced array arrangement; 0.015-144° angular range; automatic alignment

  • OPERATING TEMPERATURE

5-40°C

  • HUMIDITY

80% maximum for temperatures up to 31°C, decreasing linearly to 50% at 40°C. Non-condensing.