Description

The Malvern HPPS covers a wider application range than three conventional dynamic light scattering systems. The patented technology incorporated in the Non-Invasive Back-Scatter optics (NIBS™) enables high sensitivity and high concentration measurements over an exceptionally wide size range from 0.6nm to 6000nm diameter.

  • A higher sensitivity than a dedicated instrument for samples as dilute as 0.1ppm
  • The high concentration range of back-scatter optics up to 20 vol% to simplify sample preparation
  • A wider range of applications than covered by a classical dynamic light scattering system

Specification

Parameter / Specification

  • Size range*: 0.6nm to 6000nm hydrodynamic diameter
  • Sensitivity: 0.1mg/ml of Lysozyme monomer
  • Molecular weight range*: 103 – 107 Daltons
  • Concentration range*: 0.00001 vol% (0.1ppm) to 20 vol%
  • Measurement Technique: Dynamic light scattering
  • Patents Granted
    • German Patent 19725211
    • US Patent 6016195
    • Japanese Patent 2911877
  • Laser: He-Ne, 3.0mW,633nm
  • Product laser class: Class 1 compliant, EN 60825-1:94
  • Laser attenuation: Automatic, transmission 100% to 0.0003%
  • Laser power monitor: Silicon photodiode
  • Detector: Avalanche photodiode,Q.E > 50% at 633nm
  • Temperature control range: 8ºC to 55ºC
  • Temperature Control: Thermo-electric Peltier heater/cooler element
  • Optics purge: Input available for dry air or Nitrogen
  • Dimensions: 280mm(w) x 240mm(h) x 520mm(d)
  • Weight: 16kg
  • Power Requirements: AC 90-240V, 50/60Hz
  • Power Consumption: Max 50W
  • Ambient operating conditions: Temperature 15ºC - 40ºC
  • Humidity: 20 – 70% non-condensing
  • Operating system compatibility: Windows ’98, NT4, 2000